next up previous contents index
Next: 5.8 Quarter-Second Processing Up: 5. Calibration and Performance Previous: 5.6 Photometric Accuracy

Subsections



5.7 The Strong Source Correction

5.7.1 The reason for the correction

The responses of detectors LW1-LW4 are non-linear when they are exposed to strong sources. This non-linearity means that the photocurrent is lower than expected for the flux incident on the detectors. This occurs because the voltage produced by the radiation on the detectors de-biases them. This de-biasing lowers the responsivity hence an increase in the input signal does not produce the same increase in the output voltage, the value is lower and so the ramps are curved.

Figure 5.10: A curved ramp for a typical strong source. The green dashed lines show the change in voltage and time. The gradient of the green continous line gives the value of $\Delta V/\Delta t$. The red line indicates a guess at the gradient of the ramp without de-biasing.
\resizebox {13cm}{!}{\includegraphics{curved_ramp_lines.ps}}

Figure 5.10 shows a curved ramp for a typical strong source. The gradient of the green line shows the value of $\Delta V/\Delta t$ (see Section 4.3.6) for the ramp (the dashed lines show the change in voltage and time). The red line indicates the initial gradient of the ramp where the detector does not suffer from de-biasing. Thus, it can be seen that due to de-biasing the photocurrent obtained from the $\Delta V/\Delta t$ method underestimates the true source flux of strong sources. This becomes worse with stronger sources as the ramp curves more and could eventually flatten off. For any detector the instrument transmission varies across the bandpass filter. At the edges of the detector filters there is a low response to any signal and the effects of non-linearity are correspondingly small. Where the instrument transmission is higher the signal suffers from more non-linearity as the flux on the detector is higher. This difference across the instrument transmission means that an individual detector spectrum is more non-linear near the centre of the wavelength range, where generally the transmission is high, than it is at the short and long wavelength cut-offs. This produces strange shaped (saggy) sub-spectra (see example in Figure 6.11).

5.7.2 The correction

This non-linearity can be determined using the strong source Saturn, which is well modelled and is well calibrated in the short wavelength detectors (Davis et al. 1996, [13]). The flux model and the instrument transmission (RSRF) are used to find the photocurrents which we would expect for a linear system. These are then compared with the photocurrents found from the LWS observations of Saturn.

This comparison is shown in Figure 5.11 where the model versus LWS photocurrents ($I_{\rm model}$ and $I_{\rm data}$) are plotted as a function of wavelength. There is clearly a relationship between the two and we fit the second order polynomial (shown by the dashed line in Figure 5.11):


\begin{displaymath}
I_{\rm model}=a~ I_{\rm data} + b~ I_{\rm data}^2 \ ,
\end{displaymath} (5.5)

where $a$ and $b$ are the first and second order coefficients. This is done for all of the detectors affected by non-linearity (LW1-LW4). The coefficients are then applied to any source photocurrents ( $I_{\rm source}$) to produce the photocurrents corrected for the non-linearity ( $I_{\rm corrected}$) using

\begin{displaymath}
I_{\rm corrected}=a~ I_{\rm source} + b~ I_{\rm source}^2
\end{displaymath} (5.6)

These corrected photocurrents are then calibrated using the LCGR file derived from Uranus as usual (see Section 5.2) to produce the flux spectrum.

Figure 5.12 shows the result of this correction for the HII region W28A2 (the uncorrected spectrum is shown in Figure 6.11). To remove the scaling of individual sub-spectra that was introduced by the corrections, the sub-spectra were also scaled together using ISAP to produce a relative calibration.

Figure 5.11: Plot of the modelled versus observed photocurrents of Saturn for LW3. The dashed line shows the second order fit to the data.
\resizebox {13cm}{!}{\includegraphics{sat_mod_datao.ps}}

Figure 5.12: The plot shows part of the spectrum of the HII region W28A2 with the strong source correction applied (the uncorrected spectrum is shown in Figure 6.11). The detector sub-spectra have been scaled together to assist comparison of the shapes of the overlaps.
\resizebox {13cm}{!}{\includegraphics{w28a2_zoom_sat_col.eps}}

This shows that the sub-spectra now have similar shapes. From the figure it can be seen that the `saggyness' has been removed from the sub-spectra and they line up smoothly. Also fringing, which is seen in LWS observations of extended sources, is observed in the corrected W28A2 spectrum (fringing is discussed in Section 2.3 and in Section 6.2).

This non-linear flux response correction is applied to LW1 and LW2-LW4, i.e. one unstressed detector and three of the four stressed detectors. Table 5.8 gives the instrument responsivity for each LWS detector, as reported in Swinyard et al. 2000, [41] grouped according to detector type (also see Section 2.6 for bias voltages). The responsivity of LW1 is 3-4 times greater than that of the other unstressed Ge:Ga detectors. This high responsivity is the cause of the non-linearity and explains why this detector requires the correction. The responsivities of detectors LW2-LW4 are also high, but that of the last stressed detector, LW5, is much lower (by a factor of 3-8). Due to this low responsivity, LW5's data do not suffer noticeably from non-linearity and hence do not require the correction. Section 6.10 explains how to decide if your data need the correction, how to get them corrected and the procedure that is followed.


Table 5.8: The in-orbit detector responsivities.
Detector Type Responsivity [A/W]
     
SW1 Ge:Be 0.132
     
SW2 Ge:Ga(u) 0.070
SW3 Ge:Ga(u) 0.094
SW4 Ge:Ga(u) 0.084
SW5 Ge:Ga(u) 0.078
LW1 Ge:Ga(u) 0.274
     
LW2 Ge:Ga(s) 0.886
LW3 Ge:Ga(s) 2.365
LW4 Ge:Ga(s) 1.803
LW5 Ge:Ga(s) 0.337

5.7.3 Future

Further investigations are underway, which should result in a new way of doing the strong source correction, which will be more reliable and produce better calibrated spectra. This will correct spectra for non-linearities seen in detector sub-spectra by their saggyness (as described in Section 5.7 and 6.10). These non-linearities are due to the de-biasing of the detectors (LW1-LW4) which makes the integration ramps lower than expected, leading to an underestimation of the signal. Also the de-biasing of the detectors changes the relative spectral response of the detectors (the overall responsivity decreases and the spectral response shape changes). A new strong source correction to account for both of these should be available in a future release of LIA.


next up previous contents index
Next: 5.8 Quarter-Second Processing Up: 5. Calibration and Performance Previous: 5.6 Photometric Accuracy
ISO Handbook Volume III (LWS), Version 2.1, SAI/1999-057/Dc