The optical design of the camera (Figure 2.1) was based on an imaging lens focussing the sky image provided by the ISO telescope onto the detector. At the focal plane of the ISO telescope, a field mirror (called `Fabry mirror' in Figure 2.1) imaged the telescope exit pupil onto the camera pupil. The filters and the CVF were located at the camera pupil position. An aperture stop was placed inside the filter mounts. For the CVF, an aperture stop was located before the precise pupil location. The field mirror was also used as a field stop.
Radiation from the sky entered the camera via a pyramidal mirror and first encountered an entrance wheel offering the alternatives of a clear aperture or a set of three polarising grids with position angles 120 degrees apart; the zero orientation was defined as the spacecraft y- axis (for a description of the spacecraft axes, see the ISO Handbook Volume I, [40]). The selection between the two channels of the camera, SW and LW, was achieved by opting to use one of four tilted field mirrors mounted on the selection wheel. These field mirrors were placed in the focal plane of the telescope, and depending upon which of them was in position, the telescope beam was fed into one of the two channels.
Each channel included a filter wheel, with bandpass filters and
Circular Variable Filters (CVFs). The SW channel contained 13 filters
(including two redundant ones), and a CVF for the 2.273 to
5.122m range. The LW channel contained 10
filters, and 2 CVF's covering the 4.956 to 17.34
m ranges. The
spectral resolution was about 40 for the CVF's and ranged from 2 to
30 for the bandpass filters.
In each channel a lens, mounted on a wheel, re-imaged the focal
plane of the telescope onto the array. Four different lenses on the
wheel provided four different magnification factors matching the
fixed physical pixel size to the desired pixel field of view (pfov
hereafter)
on the sky. The pfov for the different lenses were: 1.5, 3, 6
or 12
(see Table 2.1).
Each of the field mirrors, described earlier, yielded an image of the telescope pupil. This image was located between the field mirror and the lens, in the plane of the filter wheel. Each filter carried a diaphragm, actually a hole punched through a sheet of metal, which acted as an aperture stop. For the CVF, of course, such a stop could not be implemented at this location (i.e. precisely at the filter location). Instead, a stop was placed 6 mm ahead of the plane of the pupil image, slightly oversized because of the beam aperture. The fixed filters were tilted with respect to the optical axis, to avoid ghosting due to back-reflected light. For mechanical reasons, the CVF had its plane normal to the axis. This, together with the poorer aperture stop, produced more straylight for the CVF than for the fixed filters (see Section 4.9).
The field mirror acted as a field stop for the light coming from the
telescope. For each channel, there were actually two possible field mirrors
providing two different fields of view: arcsec
, and
arcsec
.
With the 1.5
and 3
pfov, only
the central part
of the 3
diameter
unvignetted field of view provided by the
telescope beam was used by ISOCAM; the field of view of the camera
was then limited by the size of the detector array.
In the 6
pfov, there was some vignetting
in the corners of the
arcmin
square field of view covered
by the array, since the system was designed to match the 3
diameter
unvignetted field of view of the telescope. The vignetting can be up to
20% in extreme cases because of the wheel repositioning jitter (see
Section 2.2.2).
With the 12
pfov,
the field stop was always the field mirror, and
the outer part of the detector array was not illuminated.
pfov | Field mirror | Effective field of view |
1.5 ![]() |
small | ![]() |
3
![]() |
small |
![]() |
3
![]() |
large |
![]() |
6
![]() |
large |
![]() |
12
![]() |
large |
![]() |
The Point Spread Function (PSF) is a convolution of the diffraction
figure of the 60 cm telescope with the sky sampling of the pfov.
The PSF (see Section 4.6) was
undersampled with the 12
and the 6
pfov; it covered about 2 pixels for the 3
pfov,
and it was better sampled with the 1.5
pfov.
The PSF was uniform over the circular 3
field of view,
but was somewhat degraded in the corners of the square 3
field of view.
Due to the tilted field mirror, there was a small field distortion.
For the 6
pfov, an object which would have
given an image of 32
pixels on the top of the array gave an image of only about 31 pixels
at the bottom of the array (see Section 4.10).
Since it was not possible to have a broad-band anti-reflection
coating on the detector, light was reflected back from the detector.
This reflection was a source of straylight. To avoid strong ghosts the
fixed
filters were tilted with respect to the optical axis. Nevertheless,
secondary reflections on the wheels, and on the filter mounts caused
residual ghost images (see Section 4.9).
The worst case arose for the 3
pfov where
light coming from the whole 3
field of view fell on the golden
connecting strips at the edges of the photosensitive part of the
detectors. To reduce this effect in the specific case of the
3
pfov, the optional small field mirror was
provided to define a field better matched to the array size in this optical
configuration. This field mirror was undersized to take into account
positioning tolerances. Only
pixels were illuminated with this small field mirror and the
pfov lens. When the CVF was used in combination with the
3
pfov, the use of the small field mirror was mandatory.
The most frequently applied optical configurations, as recommended by the
instrument team and selected by the observers, were:
Small field mirror:
Large field mirror:
There was a certain amount of play in ISOCAM's wheels. This caused the field mirror to occasionally shift the field of view away from column 0 or 31, the edge-columns of the detector array. Column 0 was more often affected than column 31. The vignetting can be seen in LW measurements when the background is strong. The shift of the field of view also occured in the SW channel, as evidenced in the trend analysis of the CAM daily calibration measurements (Gallais & Boulade 1998, [35]).
Sources observed through the three polarisers were displaced by several
pixels on the detector array as compared to the source position
obtained through the entrance hole. The displacement was strongly
dependent upon the polariser in place. Taking the source position obtained
through the entrance hole as the
nominal (zero) position and measuring the relative displacement
of the source centre for each polariser, the results given in
Table 2.2 were obtained. For polarisation observations
using
the observing mode designed for extended sources, the displacements
of the different polarisers were compensated for by offsetting the
spacecraft pointing
(Section 3.4) in opposition to the numbers
given in Table 2.2. As a consequence of this
approach it was possible to take, as a first order approximation, the
polariser images directly obtained, and to calculate the Stokes vector
without applying further registration techniques.