The LWS diffraction grating was mounted in a scanning mechanism which rotated through , allowing to cover the extended range of wavelengths of each detector. The wavelength corresponding to each scanning position was determined by the grating angle (angle between the input aperture direction and the normal to the grating) and the detector angle (angle between the input aperture direction and the detector direction), which was a constant for a given detector.
where is the groove density of the grating (7.9 lines per mm) and is the order: the grating was used in first order for the wavelength range 84-197 m with the five long wavelength detectors LW1 to LW5, and in second order for the wavelength range 43-93 m with the five short wavelength detectors SW1 to SW5.
In operations, the grating position was actually monitored via the engineering unit called LVDT (linear variable differential transformer). Therefore, once the ten detector angles were known, the wavelength calibration consisted in finding the relationship between the engineering units LVDT and the actual grating angle .
This was done by fitting a third order polynomial to a large database consisting of the measurements of emission line centroids in terms of LVDT units associated with the expected wavelengths of the lines for a number of calibration sources observed throughout the ISO mission.
line id. | det. | LVDT | # obs | sources |
mean() | ||||
[O III]51.815 | SW1 | 1327(3) | 171 | NGC 6543 NGC 6826 G298.228 IRAS 15408 |
NGC 7027 | ||||
[O III]51.815 | SW2 | 2783(3) | 196 | NGC 6543 NGC 6826 G298.228 IRAS 15408 |
NGC 7027 NGC 6302 | ||||
[N III]57.330 | SW2 | 1993(2) | 161 | NGC 6543 NGC 6826 G298.228 IRAS 15408 |
NGC 6302 | ||||
[N III]57.330 | SW3 | 3376(5) | 152 | G298.228 IRAS 15408 NGC 6302 |
[O I]63.184 | SW2 | 1124(3) | 88 | G298.228 IRAS 15408 NGC 7027 NGC 6302 |
NGC 7023 IRAS 23133 | ||||
[O I]63.184 | SW3 | 2584(3) | 97 | G298.228 IRAS 15408 NGC 7027 NGC 6302 |
NGC 7023 IRAS 23133 | ||||
[O III]88.356 | SW5 | 1579(3) | 185 | NGC 6543 NGC 6826 G298.228 IRAS 15408 |
NGC 7027 NGC 6302 | ||||
[O III]88.356 | LW1 | 3142(4) | 189 | NGC 6543 NGC 6826 G298.228 IRAS 15408 |
NGC 7027, NGC 6302 | ||||
[N II]121.889 | LW2 | 2176(2) | 7 | NGC 6302 |
[O I]145.525 | LW3 | 1878(4) | 76 | G298.228 IRAS 15408 NGC 7027 NGC 7023 |
IRAS 23133 | ||||
[O I]145.525 | LW4 | 3250(5) | 80 | G298.228 IRAS 15408 NGC 7027 NGC 6302 |
NGC 7023 IRAS 23133 | ||||
[C II]157.741 | LW3 | 945(4) | 90 | G298.228 IRAS 15408 NGC 7027 NGC 6302 |
NGC 7023 IRAS 23133 | ||||
[C II]157.741 | LW4 | 2374(3) | 91 | G298.228 IRAS 15408 NGC 7027 NGC 6302 |
NGC 7023 IRAS 23133 |
The wavelength standards are mainly planetary nebulae and HII regions. They were chosen so as to provide the largest possible sample of lines and so that several of them were visible from ISO as much as possible during the mission (see Table 5.12 and Figure 5.22). The lines used had to be strong enough to give good signal-to-noise and to be unresolved by the grating.
|
The observations were performed weekly with end-to-end grating scans and provided measurements of seven different emission lines, spread between m and m, appearing on two detectors each.
Note that there were no measurements for SW4 and LW5, as no strong lines were found in their wavelength range. However, the relationship is in principle independent of the detector and all measurements of all lines were used together.
The wavelength calibration was first derived by adopting the detector angles measured before launch. Then the plot of the residuals (normalised differences between the expected wavelengths and the wavelengths derived from the LVDT with the polynomial relationship) showed systematic offsets for some detectors, suggesting that some of the detector angles had changed after launch. Therefore their values have been slightly adjusted until minimising the residual offsets for all detectors. The new angles used from OLP Version 6.0 onwards are listed in Table 5.13 together with the corresponding shifts relative to the pre-launch angles. The angle shifts for detectors SW4 and LW5 were adopted from the neighbouring detectors.
SW1 | SW2 | SW3 | SW4 | SW5 | LW1 | LW2 | LW3 | LW4 | LW5 | |
angle | 67.80 | 58.74 | 49.71 | 40.73 | 31.72 | 63.26 | 54.29 | 45.27 | 36.275 | 27.32 |
shift | 0.10 | 0.01 | 0.00 | 0.00 | +0.02 | +0.02 | +0.01 | +0.01 | +0.04 | +0.04 |
The stability of the system was checked by monitoring the measured LVDT at the line centres in the weekly observations. It is found to be remarkably stable for measurements performed close to the rest (central) position of the grating (LVDT2100). But elsewhere, a little jump happened in revolution 346. The jump was bigger the farther away the grating was from its rest position, and the jump had opposite signs for opposite angles (see Figure 5.23). After the jump, only a very slow drift was observed in the LVDT measurements. This jump implied that the relationship between grating angle and LVDT reading had changed on revolution 346 for an unknown reason and it was decided to derive a time-dependent wavelength calibration, which considers two distinct periods, i.e. pre- and post-revolution 346.
|
In Table 5.14 the values of the coefficients used by the pipeline are reported.
revs | 0th order | 1st order | 2nd order | 3rd order |
1-345 | 69.624422 | 5.16459527 10 | 5.02618935 10 | 8.20047303 10 |
346-875 | 69.554848 | 5.13190430 10 | 5.02794834 10 | 8.18631699 10 |
The accuracy of the grating wavelength calibration has been checked by measuring the central wavelengths of the lines observed in a large number (65) of Auto-Analysis results from observations of NGC 7027, NGC 6543, S106 and W Hya. This check has shown that in an individual observation the wavelength calibration is measured with an accuracy better than 1/4 of a resolution element (i.e. 0.07 m for SW detectors and m or LW detectors). Only in one case the errors were slightly higher for an observation performed in a revolution just preceding the jump, when the noise on the LVDT reading was the highest, but in most of the cases the wavelength determination was better than 0.1 resolution elements.
It has to be mentioned that, because of the effect illustrated in Figure 5.23, the wavelength accuracy is higher near the centre of a detector. Therefore a slight wavelength error can be observed for a line detected at a detector edge. In this case, the measurement of the line should be performed on the adjacent detector, where it is likely to fall more near the centre.
Mode | Accuracy |
grating | 25% of a resol. element |
0.07 m for SW detectors | |
0.15 m for LW detectors |