 
 
 
 
 
 
 
 
 
 
 
| Sect.   | Band | Order | Aper. | Filter   | Area   | Detector | Wavelength   | Resolution   | L  | |
| [  ] | type | number | [  m] | (  ) | AOT | |||||
| SW | 1A | 4 | 1 |   | 14-20 | InSb | 1 - 12 | 2.38 - 2.60 | 1870 - 2110 | 756 | 
| SW | 1B | 3 | 1 |   | 14-20 | InSb | 1 - 12 | 2.60 - 3.02 | 1470 - 1750 | 1043 | 
| SW | 1D | 3 | 2 |   | 14-20 | InSb | 1 - 12 | 3.02 - 3.52 | 1750 - 2150 | 1282 | 
| SW | 1E | 2 | 2 |   | 14-20 | InSb | 1 - 12 | 3.52 - 4.08 | 1290 - 1540 | 867 | 
| SW | 2A | 2 | 2 |   | 14-20 | Si:Ga | 13 - 24 | 4.08 - 5.30 | 1540 - 2130 | 2115 | 
| SW | 2B | 1 | 2 |   | 14-20 | Si:Ga | 13 - 24 | 5.30 - 7.00 | 930 - 1250 | 1377 | 
| SW | 2C | 1 | 3 |   | 14-20 | Si:Ga | 13 - 24 | 7.00 - 12.0   | 1250 - 2450 | 4276 | 
| LW | 3A | 2 | 1 |   | 14-27 | Si:As   | 25 - 36 | 12.0 - 16.5 | 1250 - 1760 | 2047 | 
| LW | 3C | 2 | 2 |   | 14-27 | Si:As   | 25 - 36 | 16.5 - 19.5 | 1760 - 2380 | 1879 | 
| LW | 3D | 1 | 2 |   | 14-27 | Si:As   | 25 - 36 | 19.5 - 27.5 | 980 - 1270 | 2524 | 
| LW | 3E | 1 | 3 |   | 20-27 | Si:As   | 25 - 36 | 27.5 - 29.0 | 1300 | 500 | 
| LW | 4 | 1 | 3 |   | 20-33 | Ge:Be | 37 - 48 | 29.0 - 45.2 | 1020 - 1630 | 4324 | 
| LW | 4A   | 2 | 1 |   | 20-33 | Ge:Be | 37 - 48 | 12.0 - 16.6 | 1250 - 1760 | 2047 | 
| LW | 4C   | 2 | 2 |   | 20-33 | Ge:Be | 37 - 48 | 16.0 - 19.7 | 1760 - 2380 | 1879 | 
| LW | 4D   | 1 | 2 |   | 20-33 | Ge:Be | 37 - 48 | 19.6 - 27.6 | 980 - 1270 | 2524 | 
| FP1 | 5A | 3 | 1 |   | 10-39 | Si:Sb | 49 - 50   | 11.4 - 12.2 | 20600 - 24000 | |
| FP1 | 5B | 2 | 1 |   | 10-39 | Si:Sb | 49 - 50   | 12.2 - 16.0 | 24000 - 32000 | |
| FP1 | 5C | 2 | 2 |   | 10-39 | Si:Sb | 49 - 50   | 16.0 - 19.0 | 32000 - 34500 | |
| FP1 | 5D | 1 | 2 |   | 10-39 | Si:Sb | 49 - 50   | 19.0 - 26.0 | 34500 - 35500 | |
| FP2 | 6 | 1 | 3   |   | 17-40 | Ge:Be | 51 - 52   | 26.0 - 44.5 | 29000 - 31000 | |
 = total number of scanner steps in band.
 = total number of scanner steps in band. m for SWS07.
m for SWS07. 
 
 
The design of the SWS instrument subdivides the 2.38-45.2 m
wavelength  range into several different bands. There are 12 grating
bands and, for the  reduced FP spectral range, 5 FP bands.  Sometimes
these bands are also referred to as AOT bands. The
bands, listed in Table 2.1, are combinations of
detector  array, aperture and grating orders such that for each band its
detector array sees a unique order of light, and
hence a unique wavelength. Therefore, a request to observe one
wavelength of light defines which array, aperture and grating order to
use. At any given time, the  astronomical source of interest to the
observer was centred on one and only  one of the three SWS
apertures, feeding light on many, or all,  of the 52
detector elements via the two rotatable scanning mirrors. While it  was
possible to find scanner settings such that all of the detector arrays 
would contain valid data, the instrument was normally operated such that
only  one or two of the detector arrays received  one order of the
grating  spectrometer, the other four or five receiving a mixture of
orders (and hence  producing no valid data). However, the observer or
archive user will receive data from all 52  detectors.
m
wavelength  range into several different bands. There are 12 grating
bands and, for the  reduced FP spectral range, 5 FP bands.  Sometimes
these bands are also referred to as AOT bands. The
bands, listed in Table 2.1, are combinations of
detector  array, aperture and grating orders such that for each band its
detector array sees a unique order of light, and
hence a unique wavelength. Therefore, a request to observe one
wavelength of light defines which array, aperture and grating order to
use. At any given time, the  astronomical source of interest to the
observer was centred on one and only  one of the three SWS
apertures, feeding light on many, or all,  of the 52
detector elements via the two rotatable scanning mirrors. While it  was
possible to find scanner settings such that all of the detector arrays 
would contain valid data, the instrument was normally operated such that
only  one or two of the detector arrays received  one order of the
grating  spectrometer, the other four or five receiving a mixture of
orders (and hence  producing no valid data). However, the observer or
archive user will receive data from all 52  detectors.
|   | 
|   | 
 
 
 
 
 
 
 
 
