Effects seen in data induced by
incorrect target coordinates or crowded fields

K. Leech, A. Heras

12 June 1997


Postscript version (compressed)

1 Introduction

This note summarizes possible effects seen in the SPD and AAR data products of SWS caused by target coordinates being incorrect by more than tex2html_wrap_inline205 4 arcsec or the target being in a crowded field.

As background, the ISO DUM, section 3.4 on `Pointing Performance and Considerations' and SWS IDUM sections 3.2.2 on `Entrance apertures and spacecraft axis', and 4.6 on `Pointing accuracy' should be consulted.

2 Description of effect

Observers may note what appears to be `noise' in their data. In the AAR this effect manifests itself as sudden flux jumps between adjacent wavelength regions and the up and down scans having different flux levels. Both these effects can be seen in figure 2.1.

   figure85
Figure 2.1: Effect of pointing as seen in AAR. Top plot is the down scan (up in wavelength), bottom plot is the up scan.

This `noise' can be seen in the SPD (see figure 2.2) as sudden jumps in the signal levels with time. A careful check of the SPD will show no signs of glitches or signal jumps (such as single detector signal jumps) that could cause this effect. The jumps will not be correlated in the up and down scans.

   figure94
Figure 2.2: `noise' as seen in the SPD. Up scan and wavelength observed as a function of time

Also, the signals (in either the SPD or AAR) can appear to be much noisier than the signals of other objects with similar flux levels.

3 Reason

Currently this effect is understood to be a function of incorrect positions being given for the observation (or the object studied being in a crowded field), the instrumental beam profile and the pointing jitter of ISO.

tex2html_wrap_inline207 As is reported in ISO INFO issue 10 (http://www.iso.vilspa.esa.es/isoinfo/isoinfo-10/) the pointing jitter was improved from tex2html_wrap_inline205 4 arcsec at the start of the mission and is now tex2html_wrap_inline205 1 arcsec.

tex2html_wrap_inline207 At the short wavelengths the instrumental beam profile has a triangular shape in the y (non-dispersion) direction and a top-hat shape in the z (dispersion) direction. A preliminary reduction of the beam profile data for band 1D (3.1 tex2html_wrap_inline215 ) shows that the beam profile goes from 100% to 10% smoothly going from the centre to tex2html_wrap_inline205 15 arcsec away in the y-direction. In the z-direction however, it goes from 90% to 10% going from 3 to 9 arcsec away - the beam profile is very steep in this direction.

If an object is misaligned from the slit centre in the y-direction then as the pointing jitters the object will move up and down a slowly varying response function. Hence the incident flux seen by the detectors will vary smoothly by small amounts. This will probably not be seen in the data.

If, however, an object is misaligned from the slit centre in the z-direction, slight shifts in the spacecrafts pointing will cause the object to be moved up and down the very steep beam profile. As an example, a misalignment of 5 arcsec in the z-direction in band 1D will reduce the incident flux on the detectors by 50%. The 1 arcsec jitter will then move it from the 70% contour to the 30% one - the incident flux can change by more than a factor of 2 on short time intervals.

Figure 3.1 shows an example of the correlation between flux jumps and pointing jitter. The observation was performed approximately 10 arcseconds off-source in the z-direction. The three figure panels cover exactly the same period of time. The top and the middle panel display the AAR, flux versus wavelength and versus time, respectively, taking only the downscan. Note that the flux values for the twelve detectors are associated by definition with the start time of the corresponding ramps, which is why in the middle figure flux values are clustered on the x-axis. The bottom panel shows the evolution of the pointing jitter dz in the dispersion direction. The vertical lines indicate the start time of each ramp (to better visualize the relation with the middle panel). As can be seen, each increasing flux jump is associated with a sharp decrease of dz to negative values during the ramp measurement, while jumps to lower fluxes are seen in relation with increases of dz to positive values.

   figure103
Figure 3.1: Correlation between AAR fluxes and spacecraft jitter

This reason for the flux jumps also explains the increased noise. Each of the detectors 24Hz readouts (contained in the ERD) are from (slightly) different incident flux levels. This introduces extra noise into the system when the pipeline tries to fit a slope to the 24Hz readout data.

4 Conclusions & Recommendations

4.1 Conclusions

This effect, primarily noticeable at short wavelengths, can happen if:

tex2html_wrap_inline207 the entered pointing is incorrect by more than tex2html_wrap_inline205 4 arcseconds

tex2html_wrap_inline207 there is another object, with a flux similar or greater than the target, 4 - 10 arcseconds away in the z-direction. This can happen in crowded fields. As noted above, since users have no way of knowing what he roll angle of the spacecraft will be it is best to ensure there are no other objects within 10 arcsec of the intended target.

Any pointing error less than tex2html_wrap_inline205 4 should not translate into this effect. Any pointing error larger than tex2html_wrap_inline205 20 arcsec (or another object being this distance away) cannot be rotated into the slit and will result in little flux falling on the detectors. As users have no way of knowing a-priori what the roll angle of the spacecraft will be for their observation, and hence do not know how a slight mispointing in RA and Dec will translate into an offset in the y and z-directions, coordinates given must be accurate to less than 4 arcsec in both RA and Dec.

4.2 Recommendations

We can only urge users to examine all entered pointings to ensure no errors have been made when entering coordinates and to go to catalogues to get the latest coordinates.

If an observer does notice such effects in their data it may be possible at a later date to correct for them, by correcting the data based on the shape of the beam profile with position as given in the IIPH.


next up previous
Up: ISO Homepage Previous: SWS Instrument Page

K. Leech, A. Heras